Bi-Dimensional Zero Padding Angular Interpolation for Arc Handling in Computed Tomography Scanner

No Thumbnail Available

Date

2015-06-14

Journal Title

Journal ISSN

Volume Title

Publisher

University of sciences and technology in Oran

Abstract

In this work we examine the accuracy of our proposed interpolation algorithm by zero-padding comparing to standard techniques for the task of interpolating additional projections to be insert in sinogram witch lost some projections caused by an X-ray tube arcing in computed tomography scanners. During the time that the X-ray tube recovers to full voltage after an arc, image data is not collected and data projections are lost caused poor image quality in the tomography reconstruction process. We have developed an algorithm based on an estimated calculation of a virtual projection using the zero-padding interpolation. The results show a significant reduction of the star effect noise on the reconstructed image. Our algorithm was compared to simple interpolation linear method using statistical hypothesis testing .Our test simulating the human body while the programming is carried out in MATLAB 6.5.

Description

Keywords

Filtered Back Projectio, Computed Tomograph, Zero Paddin, linear interpolation, Radon Projection Estimations

Citation

Collections

Endorsement

Review

Supplemented By

Referenced By